The Model 270 load cell is designed for vehicle weighing applications.
Suitable for under body installations or where a strong robust solution is required.
The Model 270 load cell is designed for vehicle weighing applications.
Suitable for under body installations or where a strong robust solution is required.
For a true zero calibration, it is necessary for the silo or vessel to be empty. However, a comprehensive health check, service, and calibration can be performed with material in the silo or vessel using the original zero calibration values. During this process, we meticulously cross-check each load cell signal and balance against previously recorded values to identify any issues such as drifting or shock loading.
Our engineers are strategically positioned across the UK to ensure prompt responses when needed. Customers with a service agreement benefit from preferential call-out rates and response times. For more information, please contact ServiceDesk@aw.co.uk
The arrow should indicate the direction from which the force will be applied. For instance, in assemblies like Compact, CleverMount, or CleanMount where the force comes from above, the arrow should point downward along the force direction. Conversely, on a floor scale where the load is applied from the foot on the floor, the arrow should point upward in the direction of the force.
Regarding products I am frequently asked if our loadcells are compatible with their existing instrumentation and vice-versa. I am also regularly asked about weighbridges and occasionally belt weigher systems.
Load cell cables can be cut to tidy the installation. However, it is advisable to ensure that each load cell in the system has cables of similar length. This helps maintain uniform load cell sensitivities and prevents potential cornering issues.

No. Each load cell in a weighing system should have the same capacity. Using load cells of different capacities will result in varied outputs among the load cells, causing weighing errors.
The calibration frequency will depend on your system and application; also, on your ISO requirements, if any. As a standard the calibration is carried out once a year, with more frequent checks on systems and more often recalibration if necessary.